Keterangan

Spectrophotometer simpler to use an higher in dependability thanks to mounting of a large-size color LCD.

High resolution satisfying European Pharmacopoeia (spectral bandpass: 1.5nm).
Trace-amount measurement in biotechnological field, etc. can be carried out using optional 50, 25 and 5micro cells.

Features
Stable optics due to double beam
In double beam design, the energy of the light source is divided into two with a half mirror so that one passes through the reference side, and the other through the sample side, which is unavailable with the single beam design.
Since the reference-side energy is also incident on a detector, photometry is carried out on the basis of this signal. Therefore, an energy change in the light source can be compensated to ensure stable measurement for a long time.
________________________________________

Incorporation of stigmatic concave diffraction grating
The optics of this instrument adopts the Seya-Namioka monochromator widespread as a representative concavediffraction grating monochromator.
Because a concave diffraction grating has both beam condensing and dispersing functions, an optical system canbe configured with fewer mirrors.
In a spectrophotometer, use of fewer mirrors signifies a shorter optical path,thus giving rise to an aberration-free bright optics.
For elimination of the aberrations which were essentially unavoidable in the past, a stigmatic concave diffractiongrating has been developed by applying Hitachi’s original technology. As a result, a higher resolution has beenrealized.

A higher resolution has been achieved by eliminating coma from the Seya- Namioka monochromator which is the most popular concave diffraction grating monochromator. Its grooving is supported by the only ruling engine in Japan. The diffraction gratings of Model U-2900/2910 have also been made with this machine.
________________________________________

Spectral bandpass 1.5nm satisfying European Pharmacopoeia
European Pharmacopoeia requires a ratio of 1.5 or larger when measuring 0.02% (V/V) solutions of toluene in hexane.

In the spectrum at right, a satisfactory value of 1.6 can be confirmed.
(Sample: 0.02% solution of toluene in n-hexane)
________________________________________

Personal computer-based instrument control and data processing.
U-2900 can be controlled by a PC equipped with optional UV Solutions Program(P/N 2J1-0310).

UV Solutions Program provides not only basic functions such as photometry quantitative analysis and wavelength scan but also the capability of data handling as described below.
(U-2910 is the model exclusively for operation with a PC.
A PC equipped with Windows®XP is required for UV Solutions Program.)

Simple to use
Buttons are laid out so as to follow the operating procedure.
Hence, operations can be performed smoothly. A rich variety of data processing functions

Spectrum can be expanded, contracted, smoothed, differentiated and integrated, and fundamental arithmetic calculations
are applicable between spectra.

Powerful support for report generation
Measured data can be sent to Microsoft® Excel. One click of the Report button suffices for data transmission.

A spectrum can be pasted to another application program.